Summary
YSM-8103 series spectrometers adopt Hamamatsu area array back-illuminated CCD (UV-sensitive or IR-sensitive version). They are very cost-effective for weak spectral measurements. The wavelength range and optical resolution are customizable; the OEM and in-depth customization are supported. YSM-8103-03 series spectrometers, equipped with Hamamatsu S16010 2048 x 64- pixel area array CCD, have higher sensitivity in 700-1100nm and are suitable for fluorescence and Raman spectral measurement.
Model List
*Other wavelength ranges and resolutions can be customized.
*Cylinder lenses can be added to some models to increase the sensitivity
Grating Selection
Slit (µm) | Grating (grooves/mm) | Wavelength Range Width (nm) | Resolution (nm) |
25 | 500 | 800 | ~1.5 |
600 | 700 | ~1 | |
1200 | 350 | ~0.5 | |
1800 | 233 | ~0.35 | |
50 | 500 | 800 | ~3 |
600 | 700 | ~2 | |
1200 | 350 | ~1 | |
1800 | 233 | ~0.7 | |
100 | 500 | 800 | ~6 |
600 | 700 | ~4 | |
1200 | 350 | ~2 | |
1800 | 233 | ~1.4 | |
200 | 500 | 800 | ~12 |
600 | 700 | ~8 | |
1200 | 350 | ~4 | |
1800 | 233 | ~2.8 |
Note: Using 1800 line grating, the width of the wavelength range in the long-wave area may not reach 233nm.
Application
Raman and fluorescence
Water and gas quality monitoring
UV-VIS weak spectral measurement